Phenomenon |
CAUSES |
Remedy |
|
No image display after turn on power supply |
Power supply Indicator not lip up |
• Power wire may disconnected from socket • Power adapter damage |
• Re-connect power wire • Replace power adapter |
Power supply Indicator lip up |
• Monitor brightness and contrast may improperly adjusted |
• Adjsut monitor brightness and contrast |
|
Probe cannot generate image on patient body or the image is unclear |
• Instrument is in Freeze status • Image control parameters, scan parameters, image processing parameters, monitoring brightness or contrast are improperly adjusted |
• Unfreeze • Properly adjust image control parameters, scan parameters, image processing parameters, monitoring brightness or contrast |
|
Probe type displayed on screen is different from actual working type |
• Poor contact of probe interface • Instrument is in Freeze status |
• In Freeze or shutdown status, re-connect probe interface • Unfreeze |
|
Black vertical bars appear on image |
• There is pollutants on probe |
• Clean probe surface |
|
There is interference noise spots or rippled interference lines on image |
•Possibly close to other equipment and magnetic field |
•Adjust B ultrasonic working environment, making it far away other equipment and magnetic field |
|
Occasionally trackball or keyboard does not work after machine is started |
• External power grid voltage is too low or has larger fluctuation. • USB equipment (U disk, USB mouse etc) is connected when starting |
•Disconnect USB equipment • Restart and shutdown machine |
|
Occassionally trackball movement is not flexible |
• Trackball is stuck with dirt |
• Clean as per method stated in Section 10.2 |